Welcome to Z&Z Medical

X-Ray Film

  • Fuji HR-T X-Ray Film Discontinued

    Fuji HRT Discontinued

    As the medical imaging community continues to migrate from analog products to digital equipment we continue to see changes in the product offerings from all of the x-ray film manufacturers. X-ray film manufacturers are making these changes to ensure they are offering the needed products for the remaining analog film users. One of the most recent changes was made by Fuji Film Corp. with the elimination of their HRT line of film. HRT was a high contrast film that catered mostly to the hospital imaging market. With the near depletion of that market the decision was made to end the manufacturing of HRT film. For customers using HRT they will need to switch to the HRU line of general purpose radiography film. Both products are green film and both offer a 400 speed system in a regular green screen. This will eliminate the need to adjust techniques when converting from HRT to HRU. The HRU will have slightly less contrast than the HRT and is slightly less expensive than the HRT. We will continue to monitor these market changes and pass them on as needed.

  • Fuji RXU X-Ray Film to be Discontinued

    Fuji Film Co. recently announced the end of production for their RXU Half Speed Blue x-ray film. If you are using the RXU x-ray film in a Blue system you can switch to the Carestream CSB.5 without any change in the speed of your system. RXU has the unique capability to work in either Green or Blue systems so Z&Z Medical sold this product as a Half Speed Blue x-ray film and as a High Speed Green x-ray film. If you were using the RXU in a Green system the Carestream CSB.5 will NOT work in your system. Z&Z Medical will not allow the return of opened boxes of x-ray film and will NOT be liable for customers trying the CSB.5 x-ray film in their Green system. Green system alternative films will be Fuji HRT, HRU or Carestream CSG. All of these products will cause your film speed to drop about 33% therefore you will need to increase your MAS 33%. If you are using the RXU film with Green Regular screens you can purchase Green Fast screens and achieve the same film speed.

    If you are unsure of the color output of your system you can call Z&Z Medical at 800-410-9575 with the name of your intensifying screens and we will determine the color output for you. You can also perform this simple test: remove the film from your cassette. Place your opened cassettes where the screen can be exposed to radiation from your x-ray equipment and you can see the screen. Set your system for a long exposure time and then turn the light down or off in the x-ray room. Expose the screen and watch for it to glow during exposure. Green systems will put off a green light and Blue systems will put off a blue light. If you have any questions or need help with this transition please feel free to contact us at 800-410-9575.

    View our selection of X-Ray Film

    View our selection of Replacement X-Ray Cassettes

  • My X-Ray Film Comes Out Black

    How often has this happened to you in the darkroom?  You take an extremely difficult x-ray, feed it into the film processor and out comes a completely or almost all black x-ray film.  You have just wasted time, money and exposed your patient to dangerous radiation with nothing to show for it.  Now you're faced with the task of determining what went wrong.

    Since x-ray radiation is the component that exposes film and turns it black, most people jump to the conclusion that the x-ray machine must be putting out too much x-ray.  This is a possibility but almost never is the cause of dark films.  There are more things that cause black films than just x-ray.  Here's a list of the most common things (in order of occurrence) that cause x-ray film to come out black or darker than normal.

    1.  Light leak in the darkroom.  This is a very common cause of dark films.  It takes a very small amount of light to expose film.  The darkroom should be checked for light leaks anytime you notice dark films.  The best way to test for a light leak is to go into the darkroom, turn out all the lights and safe light, place a new sheet of x-ray film on the counter, lay your hand on the film in the middle, and wait 2 minutes.  Process the film.  When it comes out you should not see anything but a clear film.  If you can see a dark outline around where your hand was, you have a light leak.   The best way to find where the light is entering the room is to go into the darkroom, turn out all the lights and safe light and wait 1 minute to allow the pupils in your eyes to dilate.  Now begin to look all around the room for any light that is entering the darkroom.  Pay close attention to the door on all sides, and any ceiling tiles or light fixtures.  Any amount of light that is entering the room is exposing the film when it's out of the box or cassette.  Completely seal and repair all light leaks.

    2.  Cracked red filter on the safe light.  If the red filter coating on the safe light filter gets cracked it allows white light to escape and expose film.  To test for this, go in the darkroom, turn off the white light but leave the safe light on.  Wait 1 minute and then inspect the safe light for any white light coming out of the red filter.  If any is detected, replace the filter or safe light.

    3.  Developer in the film processor is too hot.  Most automatic processors heat the developer to reduce the time it takes to develop x-ray film.  Generally most developer is heated to 92 - 95 degrees.  If the heater control circuitry malfunctions, the developer can be heated higher than this and cause overdeveloped dark films.  Check the temperature of the developer in the processor by getting a thermometer that will measure in the 90 - 100 degree range and checking the temp of the developer in the processor.  Note: Some human thermometers won't measure temperature below 95 degrees so make sure the one you use will measure that lower range.   Another cause of hot developer is the water valve got shut off or the water line was hooked to a hot water source preventing cold water from filling the wash tank which also cools the developer in some processors.  Check that cold water is coming into the processor wash tank and if not, locate the reason and repair.

    4.  Expired x-ray film All x-ray film has a life expectancy.  Most film is good for 2-3 years after it was manufactured.  The expiration date is printed on almost all boxes of film (some generic and low cost films do not print expiration dates, so buyer beware).  Check the film box for an expiration date.  It might look something like these, 01/2015,  2015 - 01, Exp Date 2015 - 01.  Expired film will develop darker than normal so dispose of any that has passed its life expectancy.

    Dark film is a common problem in x-ray.  There could be other things causing this to happen but these are the most common causes.  Give these a try and if you still have problems it might be time to call a service technician.